WHITE PAPER

Ultrafast Coherent EUV at 25-50 nm using the KMLabs XUUS4™ with the Coherent Astrella®

Xiaoshi Zhang, Eric Mountford, Matthew Kirchner, and Henry Kapteyn KMLabs, Inc., Boulder CO

The Coherent Astrella® ultrafast amplifier, combined with the KMLabs XUUS4™ high harmonic-generation system, provides stable, high quality, coherent EUV light with photon energy in the range of 25-50 eV, at flux levels enabling a range of scientific applications.

 

coherent-astrella.jpg
kmlabs-xuus4.jpg

Introduction

Just as the invention of the laser has revolutionized science and technology in recent decades, the recent development of tabletop-scale coherent laser sources at much shorter wavelengths, EUV, and soft X-rays, is likely to have a transformative impact on science and technology applications requiring laser-like performance at these short wavelengths. Coherent EUV is generated using the high-order harmonic generation (HHG) process driven by a high peak-power ultrafast laser.[1] Applications of HHG light sources include metrology for EUV lithography, spectroscopy, microscopy,[2] and dynamic studies of molecular, [3] magnetic,[4,5] material,[6] and nano systems.[7,8] Depending on application requirements, HHG source characteristics can be tailored to obtain attosecond time resolution or energy/spectral resolution as high as 30 meV/0.1 nm,[9-11] or to enable coherent imaging with spatial resolution as high as 12 nm.[2]

In the HHG process, an intense femtosecond laser is focused into a gas medium, with the high harmonics generated in the process of field ionization of the gas.[12] However, optimally implementing an HHG source is not simply a matter of focusing the light into a gas jet or cell: the efficiency of conversion into the EUV depends on multiple parameters, and can vary by orders of magnitude. In early experiments, where the objective was simply to observe the generation of high harmonics, this optimization (phase matching) was not critical. But as these sources are used in more challenging real-world applications, achieving the highest possible flux becomes increasingly critical for success. In the past, scientists often spent 1-3 years simply in implementation of a source. Even then, flux measurements in the EUV are difficult and error-prone enough that the source may be vastly un-optimized and underpowered, thus preventing success in the desired application. A successful implementation of an HHG-based EUV source requires a comprehensive approach that simplifies flux and spectrum measurement and allows for reliable EUV flux specifications.

KMLabs has developed technology to increase the efficiency of EUV generation through phasematched conversion in hollow waveguides,[13-15] and has further optimized the technique in its commercial XUUS4™ product line.[16] The three key elements of a workhorse HHG “tabletop x-ray laser” system are: (1) a robust high average power, high repetition-rate femtosecond driving laser; (2) an optimally-implemented, differentially-pumped gas target geometry for high-harmonic generation; and (3) an EUV delivery system engineered to minimize losses and manage thermal loading. This makes it possible to provide a stable, high-quality output beam for applications. KMLabs has supplied their XUUS4™ system for several years as a fully integrated and specified system with their Dragon™ and Wyvern™ ultrafast laser systems; however, the XUUS4 ™ and beamline system can be driven by other high-quality ultrafast lasers, such as the Coherent Astrella®, with appropriate pulse duration, beam quality, and stability specs. 

Figure 1

Figure 1. XUUS/Astrella system test configuration. The laser beam output from a Coherent Astrella® amplifier is telescoped to match the KMLabs XUUS4™ input laser beam parameter requirements, then is sent into the XUUS4™ system for HHG. KMLabs’ proprietary laser beam delivery system automatically aligns, stabilizes, focuses, and couples the laser beam into our patented gas-filled hollow waveguide cartridge. EUV light is generated through the laser-gas interaction and then exits into vacuum and the XUUS beamline. The beamline includes the capability for rejection of the driving laser light and for robust background-free and calibrated flux measurement and spectral characterization. The entire process is controlled and measured using KMLabs’ fully integrated XUUS4™ control software and electronics system.

 

Test Overview 

The configuration used for testing is shown in Figure 1. It consists of the Coherent Astrella®, the KMLabs XUUS4 ™, and a modularized EUV imaging spectrometer beamline which is one of several of KMLabs’ standard beamline configurations. Here we summarize test results, and show that this configuration makes it possible to produce stable EUV light output in the photon energy range up to ~50 eV (~30 nm) at flux levels that enable a broad set of scientific applications. The entire system, including the laser, the XUUS4 ™, and the beamline occupies single modest-size optical table (~5’x10’). 

 

2.1 Coherent Astrella®

The Coherent Astrella® amplifier laser system is an industrial-grade one-box ultrafast Ti:sapphire laser amplifier developed and constructed using HALT/HASS techniques for high reliability and dependable performance. It can produce up to 7 mJ per pulse at 800 nm with < 35 fs pulse width at a repetition-rate of 1 kHz (Table 1). All the laser components are in the compact (26 cm x 79 cm x 125 cm) head, including the Vitara® oscillator, stretcher, regenerative amplifier with Coherent Revolution® Q-switched Nd:YLF laser, and finally a pulse compressor. The Astrella® design has been extensively tested for reliability and robust long-term operation.

 

2.2 KMLabs XUUS4 ™

The KMLabs XUUS4 ™ is a highly engineered HHG source and beamline system that allows users to generate coherent EUV light within a day of installation. The XUUS4 ™ TM HHG source produces EUV with optimal conversion efficiency, while the beamline modules dramatically simplify the delivery of the beam to the experiment with minimum loss, maximum flexibility, and wide user customization. This approach lowers the entry barrier into this exciting and new application area of laser science and technology.

KMLabs’s XUUS4 ™ system is built on a single rugged optomechanical platform. It takes a ~0.1- 6 mJ energy, 1-100 kHz repetition-rate femtosecond (~<50 fs) laser beam as the input and, depending on the XUUS4 ™ configuration and drive laser parameters, produces a coherent EUV or soft x-ray beam with wavelength that can span from 2 to 50 nm. The KMLabs XUUS4 ™ focuses the pump light into a gas-filled hollow waveguide[13-15] that keeps the laser-focused to high intensity over an extended length, while also providing 100% overlap between the paths of the laser beam and the gas flow. The XUUS4 ™ incorporates a proprietary HHG waveguide cartridge design to implement this patented technology. Optimized differential pumping keeps gas usage to a minimum, allowing even expensive HHG gasses such as Ne, Kr, and Xe to be used economically. The EUV generated in the waveguide propagates in a vacuum path produced via efficient differential pumping to avoid re-absorption of the generated light. Compared with a gas jet, gas consumption is orders of magnitude less, and differential pumping allows for optimization of the target pressure while minimizing vacuum pump loading with any gas including Helium (this gas requires a high ~1 atm target pressure that cannot be effectively sustained using other configurations). Different gasses yield HHG spectra that are optimized for different spectral regions, with Xe/Kr ideal for longer wavelengths/lower photon energies, and He best suited for shorter wavelengths (i.e. 13.5 nm).

 

Parameter

Value

Maximum Pulse energy

7 mJ

Pulse energy used:

For 30 nm HHG

For 13 nm HHG

1.5 mJ

6 mJ

Center Wavelength 

798 nm

Nanosecond Pulse contrast at full power

1959:1 Pre

199:1 Post

Repetition Rate

1 kHz

Beam quality (M2)

1.179 X, 1.194 Y

Pulse Duration (measured by FROG)

33 fs

Beam Size (1/e2 Diameter, 2ω0)

12 mm

Power stability

<0.5% RMS for 8hrs

Table 1: Astrella® output pulse parameters as measured for optimum driving of the XUUS4™ system.

 

Since the conversion efficiency of HHG depends strongly on the peak intensity (i.e. short pulse, tightly focusable beam with M2 ~1, minimal ASE), care is also taken in delivering the optimal beam to the focal spot. The XUUS’s proprietary laser beam management system maintains both the positional and flux stability of the source over long-term operation. This is achieved thanks to a computer-assisted automatic laser beam alignment and pointing stabilization system. It also includes a convenient real-time laser beam parameter monitoring and logging system. The result is that the pointing stability of the EUV source exceeds the pointing stability performance of the driving laser— a critical requirement for reliable alignment at EUV wavelengths. The excellent long-term and short-term stability results in the most stable and usable HHG source implemented to-date. This allows the user to treat the source as a true tabletop “x-ray laser,” rather than as an experiment in itself.

 

Figure 2

Figure 2. HHG spectrum obtained from the XUUS/Astrella following multi-parameter source optimization for HHG in Argon gas. The shape of the spectrum is determined by the peak intensity of the laser, the EUV absorption of the Argon gas, and the transmission characteristics of the 200 nm aluminum filter used to eliminate the 800 nm and lower harmonics. The spectrum was found to peak at 35 nm (35 eV) with a spectral bandwidth for a single harmonic order of ~0.75 nm.The flux in the brightest 35 nm harmonic order corresponds to 1.2 x 1011 photons/second/harmonic at the source.

 

XUUS4™ Performance using Astrella drive laser

Value

Photon Flux at 35 nm (Argon) --at the source

1.2x1011 photons/sec/harmonic

Photon Flux at 13.5 nm (Helium) --at the source

1.5x107 photons/sec/harmonic

Spectral range

11 nm to 50 nm (Argon)

25 eV to 112 eV (Helium)

Beam Pointing stability

<10 µRad for 1 hour

Power stability

<6.5 % RMS for 1 hour

Table 2: Observed XUUS4™ Performance when pumped with the Coherent Astrella® amplifier.

 

Test Results

As a collaborative effort, KMLabs and Coherent conducted HHG measurements at Coherent’s facility in Santa Clara, using a production Astrella® laser and XUUS4™. After initial characterization of the Astrella® output, and implementation of a telescope to adjust the input beam size, the installation of the XUUS4™ took place over approximately one day.

We evaluated the combined XUUS/Astrella® for HHG into two wavelength regions: a) a “near EUV” spectrum generated in Argon gas with the spectrum peaked at ~35 nm; and b) a “deep EUV” spectrum generated in Helium gas with spectrum peaked at 13.5 nm. For case a) the optimal pulse energy with the Astrella® laser was found to be 1.5 mJ, while for case b) it was 6 mJ. It is important to note that the full output of the Astrella® laser is not needed for EUV generation, so a good portion of the pulse energy can be reserved for other purposes, such as pumping an OPA for pump-probe experiments.

The measured key EUV parameters are summarized in Table 2. To obtain a quantitative measurement of flux, we used an EUV-sensitive CCD camera cross-referenced to a NIST-traceable EUV photodiode, along with beamline elements with previously-measured throughput. The beamline test configuration and calibration methodology is discussed in more detail in Appendix A.

 

Figure 3

Figure 3. Stability of XUUS4™ output over 1 hour in the case for the case of HHG to 35 nm using Argon, as in figure 2. The source exhibits excellent power and beam-pointing stability.

 

The EUV spectrum obtained for case a) and b) are shown in Figure 2 and Figure 4 respectively, while the EUV beam power and pointing stability is shown in Figure 3. The shape of the EUV spectra are the result of several factors, including the peak intensity of the laser at the time during the pulse when the emission is bright and phase-matched, the re-absorption of the longer wavelengths in the Argon gas, and the transmission of the aluminum filters, which were used to reject the fundamental and lower-order harmonics. The spectrum recorded using Ar gas and shown in Figure 2 consists of odd harmonics (17-27th), peaked at the 23rd. The flux measured at the detector corresponds to a flux at the source of 1.2x1011 (+/-20%) photons/sec/harmonic for the 23rd harmonic at 35 nm (hν = 35 eV). The flux measured on the detector was 2.1x107 ph/sec, where in this case very thick, high loss Aluminum thin film filters were used to attenuate the beam to avoid detector saturation and ensure an accurate measurement.

In setting-up experiments using EUV light, the discrepancy between the flux at the source and the flux deliverable to an experiment can be very substantial. Optical losses even in a highly-optimized beamline correspond to a throughput typically in the range of 0.1% - 10%. The exact value depends on the throughput of optics used to refocus or select a specific wavelength, as well at the throughput of the thin-film filters used block any visible/NIR light (necessary when using detectors such as CCD’s that are sensitive to all wavelengths). The specifics of the actual experimental application will determine the required throughput of the beam delivery system, and must be considered in any calculations of experimental throughput and feasibility.

Figure 4 shows the spectrum obtained for case b), for high-energy HHG using helium gas. Here, the spectrum is peaked near 13.5 nm, and the flux after optimization was determined to be ~1.5 x 107 photons/sec/harmonic, at the source based on a measured flux at the detector of 5.3x104 ph/sec. Thus, HHG to 13.5 nm (92 eV) is possible using the Astrella®. However, this flux is several orders of magnitude lower than the flux at 35 nm. This mode of operation may be useful for some preliminary spectroscopic studies, and users should expect to use the XUUS4™ /Astrella® setup primarily for HHG up to hν~50 eV.

 

Figure 4

Figure 4. Generation of 13.5 nm EUV using the Astrella/XUUS for HHG in Helium gas. The observed flux optimized at a pulse energy of 6 mJ. The spectral bandwidth (FWHM) of the harmonic at 13.5 nm (59th) is ~0.5 nm. The total flux in this spectrum was measured at 1.5 x 107 photons/sec/harmonic at the source. This flux is several orders of magnitude lower than was obtained at 35 nm, making it only useful for a subset of experimental applications.

 

Summary and Conclusions

The XUUS4™ system is the culmination of more than 20 years of work by the Kapteyn/Murnane group at the University of Colorado, whose group pioneered 1) the development of ultrafast laser amplifier systems generating 20-30 fs pulses, 2) studies exploring the basic physics of the high harmonic generation process, and 3) experimental approaches for making use of HHG light in the most demanding applications such as high-fidelity EUV imaging. The test results summarized above provide users of the Coherent Astrella® a solid basis for what to expect when using their laser with the KMLabs XUUS4™.

 

References

1. A Rundquist, CG Durfee, 3rd, Z Chang, C Herne, S Backus, MM Murnane, and HC Kapteyn, "Phasematched generation of coherent soft X-rays," Science 280(5368), 1412-1415 (1998).

2. DF Gardner, M Tanksalvala, ER Shanblatt, X Zhang, BR Galloway, CL Porter, R Karl Jr, C Bevis, DE Adams, HC Kapteyn, MM Murnane, and GF Mancini, "Subwavelength coherent imaging of periodic samples using a 13.5 nm tabletop high-harmonic light source," Nat Photon 11(4), 259-263 (2017). dx.doi.org/10.1038/nphoton.2017.33.

3. E Gagnon, P Ranitovic, A Paul, CL Cocke, MM Murnane, HC Kapteyn, and AS Sandhu, "Soft x-ray driven femtosecond molecular dynamics," Science 317(5843), 1374-1378 (2007). 

4. D Rudolf, C La-O-Vorakiat, M Battiato, R Adam, JM Shaw, E Turgut, P Maldonado, S Mathias, P Grychtol, HT Nembach, TJ Silva, M Aeschlimann, HC Kapteyn, MM Murnane, CM Schneider, and PM Oppeneer, "Ultrafast magnetization enhancement in metallic multilayers driven by superdiffusive spin current," Nature Communications 3, 1037 (2012). dx.doi.org/10.1038/ncomms2029

5. S Mathias, C La-O-Vorakiat, P Grychtol, P Granitzka, E Turgut, JM Shaw, R Adam, HT Nembach, ME Siemens, S Eich, CM Schneider, TJ Silva, M Aeschlimann, MM Murnane, and HC Kapteyn, "Probing the timescale of the exchange interaction in a ferromagnetic alloy," Proceedings of the National Academy of Sciences of the United States of America 109(13), 4792-4797 (2012). dx.doi.org/10.1073/pnas.1201371109

6. C Chen, ZS Tao, A Carr, P Matyba, T Szilvasi, S Emmerich, M Piecuch, M Keller, D Zusin, S Eich, M Rollinger, WJ Youa, S Mathias, U Thumm, M Mavrikakis, M Aeschlimann, PM Oppeneer, H Kapteyn, and M Murnane, "Distinguishing attosecond electron-electron scattering and screening in transition metals," Proceedings of the National Academy of Sciences of the United States of America 114(27), E5300-E5307 (2017). dx.doi.org/10.1073/pnas.1706466114

7. KM Hoogeboom-Pot, E Turgut, JN Hernandez-Charpak, JM Shaw, HC Kapteyn, MM Murnane, and D Nardi, "Nondestructive Measurement of the Evolution of Layer-Specific Mechanical Properties in Sub-10 nm Bilayer Films," Nano Letters 16(8), 4773-4778 (2016). dx.doi.org/10.1021/acs.nanolett.6b00606.

8. KM Hoogeboom-Pot, JN Hernandez-Charpak, X Gu, TD Frazer, EH Anderson, W Chao, RW Falcone, R Yang, MM Murnane, HC Kapteyn, and D Nardi, "A new regime of nanoscale thermal transport: Collective diffusion increases dissipation efficiency," Proceedings of the National Academy of Sciences 112, 4846-4851 (2015). dx.doi.org/10.1073/pnas.1503449112.

9. S Eich, A Stange, AV Carr, J Urbancic, T Popmintchev, M Wiesenmayer, K Jansen, A Ruffing, S Jakobs, T Rohwer, S Hellmann, C Chen, P Matyba, L Kipp, K Rossnagel, M Bauer, MM Murnane, HC Kapteyn, S Mathias, and M Aeschlimann, "Time- and angle-resolved photoemission spectroscopy with optimized high-harmonic pulses using frequency-doubled Ti:Sapphire lasers," Journal of Electron Spectroscopy and Related Phenomena 195, 231-236 (2014). dx.doi.org/10.1016/j.elspec.2014.04.013.

10. D Popmintchev, C Hernandez-Garcia, F Dollar, C Mancuso, JA Perez-Hernandez, M-C Chen, A Hankla, X Gao, B Shim, AL Gaeta, M Tarazkar, DA Romanov, RJ Levis, JA Gaffney, M Foord, SB Libby, A JaronBecker, A Becker, L Plaja, MM Murnane, HC Kapteyn, and T Popmintchev, "Ultraviolet surprise: Efficient soft x-ray high-harmonic generation in multiply ionized plasmas," Science 350(6265), 1225- 1231 (2015). 

11. T Popmintchev, M-C Chen, D Popmintchev, P Arpin, S Brown, S Ališauskas, G Andriukaitis, T Balčiunas, OD Mü cke, A Pugzlys, A Baltuška, B Shim, SE Schrauth, A Gaeta, C Hernández-García, L Plaja, A Becker, A Jaron-Becker, MM Murnane, and HC Kapteyn, "Bright Coherent Ultrahigh Harmonics in the keV X-ray Regime from Mid-Infrared Femtosecond Lasers," Science 336(6086), 1287-1291 (2012). dx.doi.org/10.1126/science.1218497.

12. JL Krause, KJ Schafer, and KC Kulander, "HIGH-ORDER HARMONIC-GENERATION FROM ATOMS AND IONS IN THE HIGH-INTENSITY REGIME," Physical Review Letters 68(24), 3535-3538 (1992). dx.doi.org/10.1103/PhysRevLett.68.3535.

13. CG Durfee III, A Rundquist, HC Kapteyn, and MM Murnane, "Guided wave methods and apparatus for nonlinear frequency generation," US Patent #6,151,155 (2000).

14. T Popmintchev, D Popmintchev, MM Murnane, and H Kapteyn, "Method for phase-matched generation of coherent VUV, EUV, and x-ray light using VUV-UV-VIS lasers," US Patent #61/873,794 (Notice of Allowance, 2015).

15. TV Popmintchev, DV Popmintchev, MM Murnane, and HC Kapteyn, "Generation of VUV, EUV, and Xray Light Using VUV-UV-VIS Lasers," (Google Patents, 2017).

16. KMLabs, "XUUS™ Coherent EUV and Soft X-Ray Source" (2017), retrieved 11/12/2017, https://kmlabs.com/product/xuus/.

 

 

Appendix A: EUV flux measurement methodology

To accurately measure HHG flux, KMLabs used the beamline configuration as shown in Figure 5. There are many possible pitfalls in the accurate measurement of HHG flux-- primary among them is to block all background light from impinging on the detector, both at the fundamental driving laser wavelength of 800 nm, and at harmonic wavelengths that are within the detection range of the detector, but outside range of wavelengths that are spectrally resolved. KMLabs has developed rigorous procedures that protect against false measurements. In the most rigorous process, the aggregate flux in an HHG beam is measured using a NIST-calibrated photodiode, with the spectrum subsequently dispersed onto a CCD to allow for the flux per harmonic order to be determined.

 

Figure 5

Figure 5: Modular beamline configuration used for the EUV spectrum and flux measurments described here. 2x VBS modules reject ~95% of the 800 nm light, while reflecting ~50% of the EUV light. The VTM toroidal mirror module refocuses the light onto the detector. The filter wheels allow for metal filters to reject the remaining 800 nm light and the lower harmonic orders, while the grating module disperses the spectrum.

 

In this set of measurements, the CCD detector was cross-calibrated at KMLabs with the NISTcalibrated diode, with flux measurements determined using the CCD. Different spectral regions are calibrated using slightly different beamline configurations. For example, in the spectrum of Figure 2, the spectrum from 30-50 nm is isolated using the configuration as outlined in Table 3.

Two filters are used to avoid any 800 nm leakage through small pinholes that are inevitably present in any thin metal filter. Furthermore, since the throughput of these filters decreases with time due to oxidation, a series of three filters is employed in-line. Any two of these is sufficient to block out-of-band radiation, allowing the aggregate transmision of each filter over the spectral region of interest to be accurately measured. The dispersed spectrum is characterized with the CCD, with the flux in a single spectral peak estimated by the integrated counts measured on this detector, previously cross-referenced to the flux measured using a NIST-calibrated diode. A flux at the source can then be backed-out taking into account the beamline efficiency. For the 13.5 nm spectral measurement, the beamline parameters are given in Table 4.

 

Element

Throughput at 35 nm

VBS (2X), IR EUV beam separator module, to absorb 800 nm while reflecting EUV.

~70% each

Aluminum filters (2X), 1000 nm and 200 nm thick, to block 800 nm light as well as lower harmonic orders.

1.8% (1000 nm) 20% (200 nm)

A toroidal refocusing mirror

~70%

Diffraction grating (+1 order)

~16%

Total beamline throughput

0.02%

Total beamline throughput without metal filters

5.5%

Table 3: Beamline configuration and throughput for measurements of near EUV spectrum around 35 nm. Beamline throughput w/o metal filters is also estimated here as many experiments are not IR sensitive such as photoemission microscopy and spectroscopy.

 

Element

Throughput at 13.5 nm

VBS (2X)

~70% each

Zirconium filters (2X)

18% (200 nm)

16% (200 nm)

Toroidal refocusing mirror

~70%

Diffraction grating (+1 order)

~28%

Total beamline throughput

0.28%

Total beamline throughput without metal filters

10%

Table 4: Beamline configuration and throughput for measurements of near EUV spectrum around 13.5 nm. Beamline throughput w/o metal filters is also estimated here as many experiments are not IR sensitive such as photoemission microscopy and spectroscopy.

 