Datasheet

Components & Accessories

Medical Laser Power Delivery Fibers

Product image of Medical Laser Power Delivery Fibers

Medical Laser Power Delivery fibers include pure silica core step-index waveguides supporting a broad range of wavelengths from visible to 2.2 µm. Core sizes range from 200 µm to 550µm with numeric apertures up to 0.24. For laser surgery applications where beam homogeneity is critical, Flat-TopFA technology can be incorporated. A range of biocompatible coatings and buffers offered including polyimide, acrylate, silicone and Tefzel®. Sterilization by radiation or ETO; supported by radiation resistant pure silica core fiber and coating material choice.
This product requires a purchase of a minimum quantity of : 85

Typical Applications

  • Endoscopy
  • Dentistry
  • Ophthalmology
  • Photodynamic Therapy
  • Surgical Procedures

Features & Benefits

  • Step Index Core diameters ranging from 200 µm – 550 µm
  • High numeric apertures up to 0.24
  • Cladding diameters from 240 µm to 600 µm
  • Biocompatible coatings: polyimide, acrylate, silicone
  • Buffers include Hytrel®, Tefzel®
Parameter Products
MM-S200/240-FTB (1404842) MM-S272/300-FTB (1404843) MM-S365/400-FTB (1404844)
Optical Specifications
Operating Wavelength 375nm to 2200nm
Core NA 0.22 ± 0.02
Geometrical & Mechanical Specifications
Cladding Diameter 240.0 ± 3.0 µm 300µm ± 6µm 400µm ± 8µm
Core Diameter 200.0 ± 4.0 µm 272µm ± 10µm 365µm ± 8µm
Coating Diameter 270.0 ± 8.0 µm 330µm ± 8µm 440µm ± 10µm
Core/Clad Offset ≤ 5.00 µm
Coating Material Low Index Acrylate
Operating Temperature Range -55 °C to 85 °C
Prooftest Level ≥ 100 kpsi (0.7 GN/m²)
OH Level Low
First Buffer Diameter 335 ± 30 µm 370µm ± 30µm 520µm ± 30µm
Second Buffer Material Tefzel®: Blue