FieldMaxII-TOP
Laser Power and Energy Meter (RoHS)

For thermal, pyroelectric, and optical sensors; measures energy up to 300 pps.
Coherent Inc.
Superior Reliability & Performance
  • Lasers and Laser-based Systems
  • Laser Measurement and Control
  • Precision Optics
  • Related Accessories

  • Large, bright, backlit LCD display
  • Digital accuracy with analog-like movement for laser tuning
  • Works with thermopile, optical, and pyrolectric sensors 
  • Measures pulse energy up to 300 pps (captures every pulse) 
  • USB computer interface with complete host control capability (Windows XP or Windows 2000) 
  • Analog output with user selectable full-scale voltage (1,2, and 5V)
  • Rechargeable battery included
  • RoHS compliant

The FieldMaxII-TOP laser power and energy meter functions with a broad range of thermopile, optical and pyroelectric sensors. Together, these sensors enable FieldMaxII-TOP to measure UV, visible, and IR laser output from the nanowatt to the kilowatt range, and to work with CW and pulsed lasers with repetition rates of up to 300 Hz.

The updated FieldMax™II-TOP (RoHS) system features an enhanced PC applications software package that provides live simultaneous monitoring and control of multiple meters from a single host computer. This software provides a quick and easy way to use your PC to monitor live data and statistics,  log data to a file, plot the data in trend and histogram charts, and tune lasers with a simulated analog needle.

Host interface commands to remotely control all meter functions are also available via LabView drivers and an ActiveX DLL server component. We have included a "Getting Started VI" and our open source applications software on the CD to help you interface our meter in LabView programming. Support for multiple simultaneous meters is supported by this LabView library.

The meter also contains an analog output and rechargeable battery. This combination of features makes FieldMaxII-TOP the most powerful and versatile laser meter in its price range.  

  more...

For detailed specifications, see part number.
Name
Description
Laser power and energy meter (RoHS compliant)
Accessories
This item is the rechargeable bat ...

Specifications
Max. Measurable Power (with thermopile sensor) 30.0 kW
Max. Measurable Power (with optical sensor) 300.0 mW
Max. Measurable Energy (with pyroelectric sensor) 300.0J
Min. Measurable Power (with thermopile sensor) 10 uW
Min. Measurable Power (with optical sensor) 1 nW
Min. Measurable Energy (with pyroelectric sensor) 1 nJ
Displayable Full-Scale Power Ranges (with thermopile sensor) 3.000 mW to 30.00 kW (auto or manual ranging, selectable)
Displayable Full-Scale Power Ranges (with optical sensor) 3.000 uW to 300.0 mW (auto or manual ranging, selectable)
Displayable Full-Scale Energy Range (with pyroelectric sensor) 3.000 nJ to 300.0J (manual ranging)
Measurement Resolution 0.1% of full scale
Max. Rep. Rate (pulsed energy measurement) 300 Hz (every pulse)
Power Sampling Rate 10 Hz
Accuracy - Digital Meter ±1.0% of reading
Accuracy - System Meter accuracy + sensor accuracy